| Design Consideration
Using Ultrasonic Flow Meters
At Custody Transfer Stations
by Edgar Bowles, Section Manager, Department of Fluids Engineering, and Terry
Grimley, Principal Engineer, GRI Metering Facility, Southwest Research Institute, San
Antonio, Texas
Historically, the flow meter of choice for natural gas transmission pipeline operators in
the United States has been the orifice meter. However, over the past five years, there has
been a heightened interest in using ultrasonic flow meters for pipeline custody transfer
measurement. Driving this interest are some operational and economic advantages that
ultrasonic meters enjoy over conventional orifice meters. For instance, ultrasonic flow
meters typically have a broader flow rate range (i.e., a larger turndown ratio) than do
orifice meters. Ultrasonic meters also have no flow restriction in the meter element, no
moving parts, minimal secondary instrumentation requiring periodic re-calibration, good
measurement repeatability, and bi-directional flow capability. However, as with any
measurement technology, ultrasonic flow meters have their operational limitations, so good
engineering practice must be followed to achieve the desired results for a given flow
meter installation. In this article, we will discuss some important aspects to consider
when designing a custody transfer meter station that incorporates an ultrasonic flow
meter.
Industry Guidelines
Measurement accuracy is a primary consideration when designing a custody transfer meter
station on a gas pipeline. To ensure that the highest attainable accuracy is achieved for
a given application, meter station designers typically rely on established industry
standards or guidelines. The first recommended practice[1] for ultrasonic gas flow meters
was published by the American Gas Association (A.G.A.) Transmission Measurement Committee
in June 1998. A.G.A. Report No. 9 is fundamentally different from previous gas industry
flow meter specifications in that it is a performance-based specification that does not
include detailed mechanical installation guidelines. Report No. 9 also states that the
meter station piping configuration is to be specified by the meter manufacturer.
A.G.A. Report No. 9 specifies limits on maximum error, maximum peak-to-peak error,
repeatability, resolution, and zero-flow reading. Report No. 9 also specifies that the
measurement bias error attributed to the effect of the meter station piping configuration
must not exceed +0.3%. The performance summary specifications from A.G.A. Report No. 9 are
shown on Figure 1. In Figure 1, qmax is the maximum gas flow rate through the meter that
can be measured within the error limits, qt is the transitional gas flow rate below which
the expanded error limit is applicable, qmin is the minimum gas flow rate through the
meter that can be measured within the expanded error limit, and qI is the actual measured
gas flow rate passing through the meter under a specific set of operating conditions.
Large meters are defined by Report No. 9 as those having a nominal diameter of 12 inches
or more. Small meters are those less than 12 inches (nominal) in diameter.
From the information provided in A.G.A. Report No. 9, how does a meter station designer
know if a given ultrasonic meter installation meets the requirements of the report? More
importantly, how can a designer determine if the greatest measurement accuracy is being
achieved with a particular meter station design? There are several means available to help
answer these important questions.
Flow Field Effects
First, a meter station designer must have a fundamental understanding of how ultrasonic
meters work. Ultrasonic meters measure the gas velocity along various acoustic paths
through the cross-section of a pipe. A mathematical algorithm converts these individual
velocity measurements into a bulk flow rate. Clearly, ultrasonic meters are sensitive to
the velocity profile of the gas stream at the point of measurement. Research has shown
that distortions in the flow stream, such as velocity profile asymmetry, swirl, or the
combination of swirl and asymmetry, can produce bias errors in the flow rate
determination.[2] To only a limited degree have ultrasonic meters demonstrated an ability
to compensate or correct for measurement bias errors resulting from flow profile
distortions. (Numerous articles describing the workings of an ultrasonic meter have
appeared in this journal and others over the past several years and the reader is referred
to these for a more in-depth explanation of the measurement of gas by ultrasonic meters).
A meter station designer should, therefore, have a reasonably good idea of the velocity
profile that will be produced by a given meter station configuration. Otherwise, the
designer runs the risk that a significant measurement bias may be built into the design.
We know that the gas velocity profile shape is controlled by the piping geometry through
which the gas flows. Elbows, tees, valves, pressure regulators, and other elements in the
flow stream can create velocity profile asymmetry, swirl, or a combination of the two.
Determining the velocity profile for a given meter station configuration can be extremely
challenging and costly. There are test data in the open literature documenting the
velocity profile produced by selected piping configurations.[3] Figure 2 shows two
different installation piping configurations under recent test at the GRI Metering
Research Facility (MRF). The database of verified piping configurations is relatively
small, but growing. Meter manufacturers also have verification test data for a limited
number of meter installation configurations. Thus, one option for meter station designers
is to choose a meter station piping configuration for which the flow characteristics are
known.
Alternatively, the designer can experimentally verify the flow performance of a specific
installation configuration. This is usually a costly option. Computational modeling may
also be used to determine the flow field produced by a given meter station design.
However, computer models tend to be difficult to set up and become rather imprecise when
several bends or other obstructions (typical of most meter station configurations) are
placed in the flow stream.
If the expected velocity profile cannot be adequately characterized at the design stage,
the designer may elect to include a flow conditioner upstream of the meter. The function
of a flow conditioner is to eliminate (or at least reduce to an insignificant magnitude)
any flow distortions created by the piping and flow control elements upstream of the
ultrasonic meter. This helps ensure that the flow measurement is not biased by any flow
stream distortions. A number of flow conditioners are currently available for this
purpose. Some example flow conditioners are pictured in Figure 3.
As a word of caution when considering the inclusion of a flow conditioner as part of a
meter station installation, recent research results from the GRI MRF[2] suggest that a
given ultrasonic meter/flow conditioner combination produces unique measurement
performance. That is, the total measurement error of a given ultrasonic flow meter may
vary, depending on the configuration of the flow conditioner and upstream piping used in
combination with the meter. Thus, care should be taken when selecting a meter/flow
conditioner combination. Otherwise, measurement bias errors may be unknowingly introduced
as a result. Because of the large number of ultrasonic meter/flow conditioner
configuration combinations, it is unlikely that a universal installation guideline can
ever be established without penalizing (with long upstream pipe length requirements) those
meters and flow conditioners capable of performing acceptably well in less stringent
installations.
Meter Calibration
During the meter station design phase, attention should also be given to meter calibration
requirements. Ultrasonic meter calibration requirements are a function of the application,
but for most custody-transfer measurement, meter calibration is considered essential.
Ultrasonic meter manufacturers are usually able to deliver ultrasonic meters that perform
within the stated error limits specified on Figure 1, without the meters being flow
calibrated. However, Figure 1 allows up to +1.0% measurement error, and most custody
transfer applications demand a much smaller error limit.
Fortunately, the total measurement error associated with an ultrasonic flow meter can
usually be reduced significantly by flow calibrating the meter to eliminate the bias
errors. Figures 4 and 5 illustrate how effective flow calibration can be at eliminating
bias errors. Figure 4 is a compilation of all as found meter calibrations
performed on large meters (i.e., in this case, meters 12 to 16 inches in
diameter) at the GRI MRF over a recent 12-month period. As found means that
the meters were tested at the GRI MRF without having undergone any prior flow calibration
and with the meter setup file configured as received from the meter manufacturer. Note
that in Figure 4, very few test points lie outside the error bounds specified in A.G.A.
Report No. 9 (i.e., Figure 1). Also note, however, that many of the test points are
somewhat removed from the zero error line.
Now observe Figure 5, which is the compilation of as left meter calibrations
for the same group of meters referred to on Figure 4. As left means that each
meter was flow calibrated against the flow references at the GRI MRF (i.e., sonic nozzles)
and then the final meter calibration factor was corrected. The data shown on Figure 5 are
for the confirmation tests that verified meter performance after correction of the meter
factor. Note that the data on Figure 5 are much closer to the zero error line, indicating
that most all of the measurement bias error was eliminated.
To illustrate the economic implications of calibrating ultrasonic meters, let us study the
following hypothetical case. Transmission-grade natural gas at a line pressure of 850 psig
is flowing at a steady velocity of 50 feet per second (i.e., at about the mid-range of a
typical meter) through a 16-inch diameter ultrasonic meter. The meter was not flow
calibrated prior to installation and was delivered from the manufacturer with a -0.7%
meter bias (i.e., the specified maximum allowable in Report No. 9). If the value of the
gas being transported is approximately $2 per thousand standard cubic feet, the -0.7%
meter bias will result in a $1.8 million error in favor of the buyer over one years
time. The current cost of flow calibrating a meter of this size and capacity at a
reputable flow lab is on the order of $15,000 to $30,000. Thus, the cost of a flow
calibration to correct for the meter bias could be recovered in approximately three to six
operating days.
Two different meter calibration methodologies are referenced in A.G.A. Report No. 9. These
are Zero-Flow Verification and Flow Calibration. A Zero-Flow Verification test checks the
transit-time measurement system of an ultrasonic meter. In this test, the meter ends are
closed off and the meter body is pressurized with a gas of known composition. With no gas
flow, the meter measures the speed of sound of the test gas. The measured result is
compared to the known (theoretical) value. This test confirms the functionality of the
ultrasonic transducers and the electronic timing circuitry. This test does not confirm the
measurement performance of the ultrasonic meter under flowing conditions. The Zero-Flow
Verification is typically performed before the meter is flow calibrated initially and
before the meter is installed in the field. The test can also be performed on a periodic
basis, after the meter has been installed at the field site.
Meter performance under flowing conditions can only be verified by a Flow Calibration
test. As noted above, a flow calibration test takes into account operational and
installation effects that may adversely affect meter performance. Flow calibrations can be
performed either in-situ, using a reference test flow meter plumbed in series at the meter
station, or off site, at a flow calibration test facility. If the meter is to be
calibrated in-situ, then consideration should be given at the design stage as to all
valving and piping requirements necessary for a reference test flow meter to be installed
onsite. Although a number of pipeline operators have performed in-situ flow meter
calibrations in the past, there are no industry standards or guidelines for in-situ or
field meter proving. However, two technical references on the subject have been produced
by Park, et al.[4] and by Gallagher.[5]
Offsite flow calibration of an ultrasonic meter is, in most cases, the preferred method
for verifying meter accuracy. Worldwide, there are a number of flow laboratories capable
of providing precision flow calibrations. The calibration test conditions should
replicate, as closely as possible, the field service conditions for the meter being
calibrated. The general consensus of the user community is that the flow meter calibration
should be performed with the same meter tube piping and flow conditioner (if one is to be
used) that will be used at the field meter station. This allows measurement biases caused
by the installation configuration to be corrected for during the meter calibration.
Selection of a meter calibration laboratory also requires some forethought. Most
high-quality test flow labs have a total measurement uncertainty of about +0.2% to 0.3%
(for two standard deviations). This measurement uncertainty level has been confirmed
experimentally through inter-laboratory testing. However, each lab does have its own
individual measurement bias errors. Therefore, if an ultrasonic meter is to be calibrated
more than once, it is recommended that all of the calibrations be performed at the same
test lab. Otherwise, changes observed in the meter performance over time may be due to
variations between the test labs, rather than variability of the meter.
Other Considerations
Two other operational effects may be important and should be considered at the meter
station design stage. First, ultrasonic meters can be adversely affected by nearby sources
of ultrasonic noise (i.e., energy emitted in the frequency range between approximately 50
to 500 kilohertz). Such noise sources may include quiet valves designed to
reduce audible flow noise, pressure regulators, and other significant flow restrictions in
the pipeline. Meter manufacturers are actively pursuing solutions to this problem and
meter station designers should consult the manufacturers for assistance. Field experience
has shown that it is best to locate the meter upstream of potential noise sources. In
addition, placing pipe bends between the meter and the noise source helps attenuate the
extraneous ultrasonic noise.
In addition, the cleanliness of the gas stream can affect ultrasonic meter performance.
Buildup (from compressor oil, condensate, or other sources) on the face of an ultrasonic
transducer can prevent the unit from transmitting ultrasonic pulses through the gas
stream. This can lead to what is called ultrasonic path dropout. Path dropout
can add to measurement error. Also, dispersed liquids in the gas flow stream (e.g., mists
or aerosols) can attenuate the ultrasonic signals and completely incapacitate the meter.
Thus, the cleanliness of the gas stream and its effects on the interior of the meter are
critical to good meter performance.
Conclusions
Acceptance of ultrasonic flow meters in natural gas transmission pipelines is growing
rapidly worldwide. By using good engineering judgment and careful design practices, users
of ultrasonic meters can realize the benefits of this new technology. P&GJ |